SmartSpectra SDK

Getting a Good Measurement

Camera setup, lighting, and positioning guidance for accurate SmartSpectra readings, what good vs. bad measurements look like, and how to read the SDK's live validation feedback.

Important: SDK metrics are offered for general wellness and informational purposes only. SDK metrics have not been cleared by the FDA and may not be used for medical diagnosis or treatment.

SmartSpectra estimates pulse, breathing, and related vitals from subtle, per-frame changes in the camera image: small color shifts in skin caused by blood volume pulse, and small movements of the chest and abdomen caused by breathing. Because the signal itself is subtle, capture conditions matter as much as the model — a well-lit, stable, centered shot with an unobstructed face (and chest, for breathing) will consistently produce high-confidence results, while poor lighting, motion, or framing will not.

This page is the practical setup guide: what to tell your users, what the SDK tells you back in real time, and what a good vs. a bad capture looks like. For the precise per-metric operating ranges and known limitations, see Model Cards and Limitations.

Quick checklist

Point end users at this before their first measurement:

ConditionRequirement
LightingEven, diffuse light on the face. Not too dark, not too bright, no harsh shadows or flicker.
FramingOne face, centered, roughly facing the camera. For breathing/chest metrics, the chest is also visible.
DistanceClose enough that the face isn't tiny in frame, far enough that it isn't clipped or overexposed.
StabilityCamera itself must stay still for breathing/chest metrics. Handheld is fine for pulse, HRV, and blood pressure waveform only.
Subject motionStay still. Avoid talking, chewing gum, or large head/body movement during capture.
ClothingAvoid all-dark clothing or tight, high-contrast stripes if measuring breathing.
PatienceSome metrics need a full window before they're trustworthy — see Confidence and warm-up time below.

A good SmartSpectra capture setup: centered, evenly lit, camera on a stable stand, face and chest in frame

Set up the environment

Lighting

Good lighting is the single biggest lever for measurement quality, because the pulse signal is a very small change in skin color.

Do:

  • Use even, diffuse light on the face — natural window light or soft indoor lighting both work well.
  • Light the face from the front or slightly off-axis, not from directly behind.
  • Keep the light source constant. Avoid sources that visibly flicker, such as some fluorescent lights, or a TV/monitor providing the primary light on the face.

Avoid:

  • Too dark. Underexposed frames don't carry enough signal, and low light also encourages some cameras to automatically lower their frame rate to lengthen exposure — which can trigger a frame-rate warning even when the picture "looks" acceptable.
  • Too bright / backlit. A window or bright light directly behind the subject silhouettes the face and blows out the parts of the frame the model needs. Overhead-only lighting that casts hard shadows across half the face has the same effect.
  • Uneven lighting. One side of the face brightly lit and the other in shadow reduces signal quality even when the average exposure looks fine.

Good lighting: evenly lit face with no shadows. Bad lighting: one-sided light casting a hard shadow across the face

Framing and distance

  • Exactly one face should be in frame. Multiple faces, or no face at all, both block processing.
  • The face should be roughly orthogonal to the camera — a face turned far to the side or tilted away doesn't track reliably.
  • The face shouldn't be so close that it's clipped or fills the whole frame, or so far away that it's a small fraction of the image.
  • For breathing and chest-based metrics, the upper chest also needs to be visible and unobstructed; the lower abdominal breathing waveform additionally needs the waistline visible.

Good distance: centered face and chest with comfortable margin. Too close: face fills and clips the frame. Too far: face is a small fraction of the frame

Camera stability

Stability requirements differ by metric:

  • Pulse rate, HRV, and the arterial pressure waveform tolerate a handheld camera as long as it's relatively stable — small hand shake is fine.
  • Breathing rate and the chest/abdomen waveforms need the camera itself to be stationary (e.g., propped on a stand, table, or dock). Handheld capture defeats breathing measurement because chest-motion tracking can't distinguish camera motion from breathing motion.

If your app supports both, default to a stable mount so the same session can measure everything requested.

Subject stillness and behavior

  • The subject should sit still — large head, body, or camera motion produces unreliable readings with deceptively high confidence, so it's worth coaching users explicitly rather than relying on the SDK to catch every case.
  • Avoid talking during a breathing measurement; speech changes chest/abdomen motion in ways that don't reflect the breathing signal.
  • Avoid chewing gum during pulse, HRV, or blood-pressure-waveform measurement.

Clothing (breathing)

Dark clothing absorbs light needed to track subtle chest motion, and tight, high-contrast striped patterns can alias against the camera's pixel grid and get misread as motion. Prefer plain, lighter-colored clothing for breathing measurements.

Frame rate

The on-device pipeline needs a sustained camera frame rate — don't let the camera fall below roughly 25 fps; target 30 fps for headroom. This is usually automatic on modern devices, but two things can undermine it:

  • Low light, which can cause the camera's auto-exposure to lengthen exposure time and drop frame rate to compensate — another reason to fix lighting first if you see intermittent frame-rate warnings.
  • Heavy on-device CPU/GPU contention from other work happening at the same time as capture.

Confidence and warm-up time

Every rate-based metric ships with a confidence value (0–100) and a stable flag alongside the value itself — treat both as part of the reading, not just the number:

  • Confidence starts at 0 and only becomes meaningful once the metric's analysis window has filled: 30 seconds for breathing rate and the breathing waveforms, 60 seconds for HRV, 12 seconds for pulse rate. Don't surface or act on a reading before its window has elapsed — early "0 confidence" samples are expected, not a bug.
  • stable indicates whether the underlying detection (face, landmarks, etc.) is currently reliable frame-to-frame. Treat stable: false the same way you'd treat a validation warning: something in the capture (framing, motion, lighting) is currently degraded.

See Data Types for the full payload schema.

Reading the SDK's live feedback

The SDK continuously evaluates the incoming frames and reports a ValidationStatus — a stable code plus a human-readable hint string — so you don't have to guess why a measurement looks bad. Surface the hint text directly to end users; it's designed to be shown as-is.

CodeMeaningWhat to tell the user
OKFrame passes all checks.Nothing — this is the "good" state.
NO_FACE_FOUNDNo face detected in frame.Get a face into frame, facing the camera.
MULTIPLE_FACES_FOUNDMore than one face detected.Make sure only one person is in frame.
FACE_NOT_CENTEREDFace detected but off-center horizontally (left/right).Center your face in the frame.
TOO_DARKImage is underexposed.Add light, or move to a brighter area.
TOO_BRIGHTImage is overexposed.Reduce light, or move away from a bright/backlit source.
CHEST_NOT_VISIBLEChest not visible or too far away (breathing/chest metrics).Reframe so the upper chest is visible.
CAMERA_TUNINGCamera is still auto-adjusting exposure/focus.Wait a moment before recording.
FRAME_RATE_TOO_LOWSustained camera frame rate has dropped too low.Improve lighting (see above) or reduce concurrent device load.
EXCESSIVE_MOTIONBody or head motion exceeds the reliable-measurement threshold.Hold still — motion may affect accuracy.
FACE_TOO_CLOSEFace fills too much of the frame.Move away from the camera.
FACE_TOO_FARFace is smaller than the minimum usable size.Move closer to the camera.
FACE_TOO_HIGHFace is too high in the frame (off-center vertically).Move down, or tilt the camera up.
FACE_TOO_LOWFace is too low in the frame (off-center vertically).Move up, or tilt the camera down.
FACE_NOT_FORWARDFace has turned or tilted too far away from the camera.Face the camera.

This is the same enum across platforms — Swift's ValidationCode, Kotlin's ValidationCode, C++'s spectra::ValidationCode, and the Node client's ValidationCodeValue all carry the same codes and wire values.

Swift:

// sdk.validationStatus is an @Observable property: (code, hint)
guard let validationStatus = sdk.validationStatus else { return }
statusLabel.text = validationStatus.hint

Kotlin:

sdk.validationStatus.observe(this) { status ->
    validationLabel.text = status?.hint ?: "--"
}

C++:

sdk.SetOnValidationStatusChanged(
    [](const spectra::ValidationStatus& status, int64_t timestamp_us) {
      std::cerr << "Validation: " << status.hint << "\n";
    });

TypeScript:

sdk.on('validationStatus', (code, timestampUs, hint) => {
  showBanner(hint);
});

Tip: Debounce or coalesce repeated identical statuses before showing them — a stationary, well-positioned subject can still flicker between OK and a borderline code frame-to-frame. Update the UI only when the code actually changes.

What good vs. bad looks like

Good capture: a clean, periodic waveform with rising confidence. Bad capture: an irregular, motion-corrupted waveform with low confidence

A good pulse or breathing trace looks like a clean, regular, periodic waveform — smoothly repeating peaks roughly in line with a plausible heart or breathing rate, with confidence rising toward 100 as the analysis window fills. A bad trace is irregular and jagged, with no consistent periodicity, confidence that stays low or drops out, and stable: false or a non-OK ValidationStatus during the affected stretch. If a trace looks noisy, check the underlying cause first — camera motion, subject motion, poor lighting, or a partially occluded face/chest — rather than assuming the reading is a valid-but-unusual result. See Model Cards and Limitations for the exact operating range and "works when / does not work when" conditions for each metric.

Guidance to give your end users

The setup guidance above translates directly into a short in-app tutorial. A concrete example (adapted from the SmartSpectra sample apps):

  1. Place your device on a stable surface, like a table or stand.
  2. Use a well-lit environment — natural daylight works best.
  3. Make sure your face is evenly lit with no strong shadows.
  4. Avoid bright light sources directly behind you, such as a window or overhead light.
  5. Stay still and avoid talking during the measurement.
  6. Watch for on-screen feedback — it tells you exactly what to fix if something's off.
  7. Start recording when prompted, and follow any auto-restart prompts if feedback appears mid-measurement.
  • Model Cards and Limitations — per-metric valid ranges, "works when" / "does not work when" conditions, and links to the full model cards.
  • Data Types — the Measurement, MeasurementWithConfidence, and ValidationStatus payload schema.